中国自然资源航空物探遥感中心主办
地质出版社出版

石墨炉原吸收光谱法测定化探样品中超痕量铂钯

官东超. 石墨炉原吸收光谱法测定化探样品中超痕量铂钯[J]. 物探与化探, 2018, (5): 1054-1058. doi: 10.11720/wtyht.2018.1362
引用本文: 官东超. 石墨炉原吸收光谱法测定化探样品中超痕量铂钯[J]. 物探与化探, 2018, (5): 1054-1058. doi: 10.11720/wtyht.2018.1362
GUAN Dong-Chao. The application of graphite furnace atomic absorption spectrometry to determination of ultra-trace platinum and ultra-trace palladium in rock ore[J]. Geophysical and Geochemical Exploration, 2018, (5): 1054-1058. doi: 10.11720/wtyht.2018.1362
Citation: GUAN Dong-Chao. The application of graphite furnace atomic absorption spectrometry to determination of ultra-trace platinum and ultra-trace palladium in rock ore[J]. Geophysical and Geochemical Exploration, 2018, (5): 1054-1058. doi: 10.11720/wtyht.2018.1362

石墨炉原吸收光谱法测定化探样品中超痕量铂钯

详细信息
  • 中图分类号: P632

The application of graphite furnace atomic absorption spectrometry to determination of ultra-trace platinum and ultra-trace palladium in rock ore

  • 研究了苯基硫脲—正辛醇体系萃取石墨炉原子吸收光谱法测定化探样品中超痕量铂钯的方法.确定了仪器的最佳条件,并从介质酸度、萃取时间和体积、苯基硫脲的浓度、萃取剂体积和干扰情况等方面进行了详细的考察.样品经王水溶解,液相萃取,以有机相进样测定.铂、钯检出限分别为0.5×10-9和0.1×10-9.对含铂14.7×10-9和钯15.2×10-9的矿样重复取样溶矿测定8次,其相对标准偏差分别为6.0%和4.7%,测试质量较好.
  • 加载中
  • 加载中
计量
  • 文章访问数:  332
  • PDF下载数:  14
  • 施引文献:  0
出版历程
刊出日期:  2018-10-10

目录